A word on 'lead': electrode, lead wire, lead
An electrode is a contact on the body. A lead is a measurement of the voltage difference between a positive input and a reference. The reference can combine signals from more than one electrode. Ashley and Niebauer (2004)
Why the electrodes are colour coded
The labels and colours on the lead wires help identify the intended electrode connections. Trace each wire to its labelled position. Crawford and Doherty (2010)
Two colour systems


The AHA and IEC conventions use different labels and colours. Identify the convention on the equipment before matching the wires to the diagram. Crawford and Doherty (2010)
AHA: American Heart Association

These are the AHA limb identifiers and colours shown in the illustration. Crawford and Doherty (2010)
- Right arm lead: ID RA, white
- Left arm lead: ID LA, black
- Right leg lead: ID RL, green
- Left leg lead: ID LL, red
- First chest lead: ID V1, red
- Second chest lead: ID V2, yellow
- Third chest lead: ID V3, green
- Fourth chest lead: ID V4, blue
- Fifth chest lead: ID V5, orange
- Sixth chest lead: ID V6, purple
IEC: International Electrotechnical Commission

These are the IEC limb identifiers and colours shown in the illustration. Crawford and Doherty (2010)
- Right arm lead: ID R, red
- Left arm lead: ID L, yellow
- Right leg lead: ID N, black
- Left leg lead: ID F, green
- First chest lead: ID C1, red
- Second chest lead: ID C2, yellow
- Third chest lead: ID C3, green
- Fourth chest lead: ID C4, brown
- Fifth chest lead: ID C5, black
- Sixth chest lead: ID C6, purple
Check the machine before you start
Check the printed electrode identifiers against the equipment’s lead set. Colour alone can be ambiguous between conventions. Crawford and Doherty (2010)