3D ECG Leads

Electrode identification

ECG electrode colour coding: AHA and IEC

Compare the AHA and IEC colours and abbreviations for each of the ten electrodes.

Electrode Placement Overview: This is the electrode placement overview. There are 4 limb electrodes (one for each limb) and 6 precordial (chest) electrodes.
A word on 'lead': electrode, lead wire, lead

An electrode is a contact on the body. A lead is a measurement of the voltage difference between a positive input and a reference. The reference can combine signals from more than one electrode. Ashley and Niebauer (2004)

Why the electrodes are colour coded

The labels and colours on the lead wires help identify the intended electrode connections. Trace each wire to its labelled position. Crawford and Doherty (2010)

Two colour systems

Electrode Placement Overview: This is the electrode placement overview. There are 4 limb electrodes (one for each limb) and 6 precordial (chest) electrodes.
Electrode Placement Overview. IEC labels and colours.Screenshot from the 3D ECG Leads app.
Ten electrodes in the AHA and IEC colour conventions.
Ten electrodes in the AHA and IEC colour conventions. (Figure 1 of the placement guide.)Illustration from the original 3D ECG Leads placement guide.

The AHA and IEC conventions use different labels and colours. Identify the convention on the equipment before matching the wires to the diagram. Crawford and Doherty (2010)

AHA: American Heart Association

Legacy figure 2 with the AHA limb labels corrected: a body with ten electrodes and the AHA legend RA, LA, RL, LL, V1 to V6
The American Heart Association ( AHA ) Color Coding System. Corrected on 23 November 2026: the AHA limb labels read RA, LA, RL, LL. (Figure 2 of the placement guide.)Illustration from the original 3D ECG Leads placement guide.

These are the AHA limb identifiers and colours shown in the illustration. Crawford and Doherty (2010)

  • Right arm lead: ID RA, white
  • Left arm lead: ID LA, black
  • Right leg lead: ID RL, green
  • Left leg lead: ID LL, red
  • First chest lead: ID V1, red
  • Second chest lead: ID V2, yellow
  • Third chest lead: ID V3, green
  • Fourth chest lead: ID V4, blue
  • Fifth chest lead: ID V5, orange
  • Sixth chest lead: ID V6, purple

IEC: International Electrotechnical Commission

12 lead ecg placement color coded
The International Electrotechnical Commission Color Coding System (IEC) (Figure 3 of the placement guide.)Illustration from the original 3D ECG Leads placement guide.

These are the IEC limb identifiers and colours shown in the illustration. Crawford and Doherty (2010)

  • Right arm lead: ID R, red
  • Left arm lead: ID L, yellow
  • Right leg lead: ID N, black
  • Left leg lead: ID F, green
  • First chest lead: ID C1, red
  • Second chest lead: ID C2, yellow
  • Third chest lead: ID C3, green
  • Fourth chest lead: ID C4, brown
  • Fifth chest lead: ID C5, black
  • Sixth chest lead: ID C6, purple

Check the machine before you start

Check the printed electrode identifiers against the equipment’s lead set. Colour alone can be ambiguous between conventions. Crawford and Doherty (2010)