Limb electrodes: standard and modified placement

Explore the standard limb electrode positions, modified placement and how placement is documented on a tracing.

Example limb electrode positions near the wrists and ankles.
A word on 'lead': electrode, lead wire, lead

An electrode is a contact on the body. A lead is a measurement of the voltage difference between a positive input and a reference. The reference can combine signals from more than one electrode. Ashley and Niebauer (2004)

Standard placement

Example limb electrode positions near the wrists and ankles.
Example limb electrode positions near the wrists and ankles. (Figure 4 of the placement guide.)Illustration from the original 3D ECG Leads placement guide.

For a standard resting recording, place the four limb electrodes on the arms and legs, distal to the shoulders and hips. The diagram shows positions near the wrists and ankles. Kligfield et al. (2007)

Where each limb lead goes

RA or R is the right arm, LA or L the left arm, RL or N the right leg and LL or F the left leg. Kligfield et al. (2007), Crawford and Doherty (2010)

When a wrist or ankle cannot be used

The illustrated locations near the wrists and ankles are examples of limb placement. Record the actual sites if a modification is used. Kligfield et al. (2007), Jowett et al. (2005)

Practical points from the app

Practical points from the app
Figure
Left Arm Electrode Placement: The Left Arm Electrode LA(L) is placed above the left wrist if possible. It can be placed on the left shoulder, if done so the right arm electrode should also be placed on the right shoulder rather than the right wrist. The arm should be resting on a surface to avoid muscle tension.
Left Arm Electrode Placement.Screenshot from the 3D ECG Leads app.
Right Arm Electrode Placement: The Right Arm Electrode RA(R) is placed above the right wrist if possible. It can be placed on the right shoulder, if done so the left arm electrode should also be placed on the left shoulder rather than the left wrist. The arm should be resting on a surface to avoid muscle tension.
Right Arm Electrode Placement.Screenshot from the 3D ECG Leads app.
Left Leg Electrode Placement: The Left Leg Electrode LL(F) is placed above the left ankle if possible. It can be placed on the left thigh, if done so the right leg electrode should also be placed on the right thigh rather than the right lower leg. The leg should be resting on a surface to avoid muscle tension.
Left Leg Electrode Placement.Screenshot from the 3D ECG Leads app.
Right Leg Electrode Placement: The Right Leg Electrode RL(N) is placed above the right ankle if possible. It can be placed on the right thigh, if done so the left leg electrode should also be placed on the left thigh rather than the left lower leg. The leg should be resting on a surface to avoid muscle tension.
Right Leg Electrode Placement.Screenshot from the 3D ECG Leads app.

Compare the app’s labelled right and left limb electrodes with the standard positions above. Kligfield et al. (2007)

Modified placement

Example limb electrode positions near the wrists and ankles.
Example limb electrode positions near the wrists and ankles. (Figure 4 of the placement guide.)Illustration from the original 3D ECG Leads placement guide.
The illustrated modified arrangement places limb electrodes on the torso.
The illustrated modified arrangement places limb electrodes on the torso. (Figure 5 of the placement guide.)Illustration from the original 3D ECG Leads placement guide.

Moving limb electrodes onto the torso can change the recorded waveforms. Document a modified placement on the recording so it can be taken into account when tracings are compared. Jowett et al. (2005)

What the modified placement costs

Bipolar Lead I: Lead I is called a Bipolar Lead. It is generated by 2 physical electrodes: LA(L) - the Exploring Electrode (+) and RA(R) - the Indifferent Electrode (-).
Bipolar Lead I.Screenshot from the 3D ECG Leads app.
Einthoven’s triangle and the three bipolar limb leads.
Einthoven’s triangle and the three bipolar limb leads. (Figure 6 of the placement guide.)Illustration from the original 3D ECG Leads placement guide.

A bipolar lead measures the voltage difference between two electrodes. Lead I compares the left arm (+) with the right arm (−), lead II the left leg (+) with the right arm (−), and lead III the left leg (+) with the left arm (−). These three views form the sides of Einthoven’s triangle in the frontal plane. Kligfield et al. (2007)

When limb electrodes move onto the torso, they sample the signal at different sites. In paired recordings, Jowett and colleagues found changes in frontal axis and waveform appearance. Record the modified sites so a later comparison is meaningful. Jowett et al. (2005)

For the lead directions themselves, see bipolar leads I, II and III.

Recommendation: use the standard positions and label any change

Document the positions used for a modified recording. Keep that information with the tracing. Jowett et al. (2005)

References

  • Kligfield et al. (2007). Recommendations for the standardization and interpretation of the electrocardiogram, Part I. JACC 49, 1109–1127.
  • Crawford and Doherty (2010). Ten steps to recording a standard 12-lead ECG. Practice Nursing 21, 622–630.
  • Jowett et al. (2005). Modified electrode placement must be recorded when performing 12-lead electrocardiograms. Postgraduate Medical Journal 81, 122–125.
  • Ashley and Niebauer (2004). Conquering the ECG. In Cardiology Explained, chapter 3.