A word on 'lead': electrode, lead wire, lead
An electrode is a contact on the body. A lead is a measurement of the voltage difference between a positive input and a reference. The reference can combine signals from more than one electrode. Ashley and Niebauer (2004)
Standard placement

For a standard resting recording, place the four limb electrodes on the arms and legs, distal to the shoulders and hips. The diagram shows positions near the wrists and ankles. Kligfield et al. (2007)
Where each limb lead goes
RA or R is the right arm, LA or L the left arm, RL or N the right leg and LL or F the left leg. Kligfield et al. (2007), Crawford and Doherty (2010)
When a wrist or ankle cannot be used
The illustrated locations near the wrists and ankles are examples of limb placement. Record the actual sites if a modification is used. Kligfield et al. (2007), Jowett et al. (2005)
Practical points from the app
Compare the app’s labelled right and left limb electrodes with the standard positions above. Kligfield et al. (2007)
Modified placement


Moving limb electrodes onto the torso can change the recorded waveforms. Document a modified placement on the recording so it can be taken into account when tracings are compared. Jowett et al. (2005)
What the modified placement costs


A bipolar lead measures the voltage difference between two electrodes. Lead I compares the left arm (+) with the right arm (−), lead II the left leg (+) with the right arm (−), and lead III the left leg (+) with the left arm (−). These three views form the sides of Einthoven’s triangle in the frontal plane. Kligfield et al. (2007)
When limb electrodes move onto the torso, they sample the signal at different sites. In paired recordings, Jowett and colleagues found changes in frontal axis and waveform appearance. Record the modified sites so a later comparison is meaningful. Jowett et al. (2005)
For the lead directions themselves, see bipolar leads I, II and III.
Recommendation: use the standard positions and label any change
Document the positions used for a modified recording. Keep that information with the tracing. Jowett et al. (2005)
References
- Kligfield et al. (2007). Recommendations for the standardization and interpretation of the electrocardiogram, Part I. JACC 49, 1109–1127.
- Crawford and Doherty (2010). Ten steps to recording a standard 12-lead ECG. Practice Nursing 21, 622–630.
- Jowett et al. (2005). Modified electrode placement must be recorded when performing 12-lead electrocardiograms. Postgraduate Medical Journal 81, 122–125.
- Ashley and Niebauer (2004). Conquering the ECG. In Cardiology Explained, chapter 3.