Why ten electrodes give twelve leads

See the difference between an electrode and a lead, and how ten contacts produce twelve views of the heart.

Einthoven’s triangle and the three bipolar limb leads.

Why twelve leads?

Why 12 Leads?: Because 3 Limb Electrodes ( RA(R), LA(L) and LL(F) ) will generate 6 Frontal Leads. 3 Bipolar Leads ( I, II, III ) 3 Unipolar Leads ( aVL, aVF, aVR ) The RL(N) is the grounding
Why 12 Leads?.Screenshot from the 3D ECG Leads app.

A standard twelve-lead ECG uses ten electrodes. The limb electrodes provide six frontal-plane leads, and six chest electrodes provide six horizontal-plane leads. Kligfield et al. (2007)

What an ECG lead is

What is a an ECG Lead ?: An ECG Lead is a trace of the heart's electrical activity, seen from a specific angle. It uses 2 electrodes to generate the reading: an Indifferent Electrode (-) and an Exploring Electrode (+). Press play for more
What is a an ECG Lead ?.Screenshot from the 3D ECG Leads app.

An electrode is a contact on the body. A lead is a measurement of the voltage difference between a positive input and a reference. The reference can combine signals from more than one electrode. Kligfield et al. (2007)

Bipolar leads I, II and III

Bipolar leads I, II and III
Figure
Bipolar Lead I: Lead I is called a Bipolar Lead. It is generated by 2 physical electrodes: LA(L) - the Exploring Electrode (+) and RA(R) - the Indifferent Electrode (-).
Bipolar Lead I.Screenshot from the 3D ECG Leads app.
Bipolar Lead II: Lead II is called a Bipolar Lead. It is generated by 2 physical electrodes: LL(F) - the Exploring Electrode (+) and RA(R) - the Indifferent Electrode (-).
Bipolar Lead II.Screenshot from the 3D ECG Leads app.
Bipolar Lead III: Lead III is called a Bipolar Lead. It is generated by 2 physical electrodes. LL(F) - the Exploring Electrode (+) and LA(L) - the Indifferent Electrode (-).
Bipolar Lead III.Screenshot from the 3D ECG Leads app.

Lead I compares the left arm with the right arm. Lead II compares the left leg with the right arm. Lead III compares the left leg with the left arm. In each pair the first limb is the positive input. Kligfield et al. (2007)

Augmented leads aVL, aVR and aVF

Augmented leads aVL, aVR and aVF
Figure
Augmented Lead aVL: Lead aVL is called an Augmented Lead. It is generated by one physical Exploring Electrode (+) represented by LA(L), and a virtual Indifferent Electrode (-) formed by RA(R) and LL(F).
Augmented Lead aVL.Screenshot from the 3D ECG Leads app.
Augmented Lead aVR: Lead aVR is called an Augmented Lead. It is generated by one physical Exploring Electrode (+) represented by RA(R), and a virtual Indifferent Electrode (-) formed by LA(L) and LL(F).
Augmented Lead aVR.Screenshot from the 3D ECG Leads app.
Augmented Lead aVF: Lead aVF is called an Augmented Lead. It is generated by one physical Exploring Electrode (+) represented by LL(F), and a virtual Indifferent Electrode (-) formed by LA(L) and RA(R).
Augmented Lead aVF.Screenshot from the 3D ECG Leads app.

For aVR, aVL and aVF, the positive electrode is on the right arm, left arm and left leg respectively. The reference is the average of the other two measuring limb electrodes. Kligfield et al. (2007)

The angle of each frontal lead

In the conventional frontal-plane diagram, I is 0°, II +60°, III +120°, aVF +90°, aVL −30° and aVR −150°. Ashley and Niebauer (2004)