Why twelve leads?

A standard twelve-lead ECG uses ten electrodes. The limb electrodes provide six frontal-plane leads, and six chest electrodes provide six horizontal-plane leads. Kligfield et al. (2007)
What an ECG lead is

An electrode is a contact on the body. A lead is a measurement of the voltage difference between a positive input and a reference. The reference can combine signals from more than one electrode. Kligfield et al. (2007)
Bipolar leads I, II and III
Lead I compares the left arm with the right arm. Lead II compares the left leg with the right arm. Lead III compares the left leg with the left arm. In each pair the first limb is the positive input. Kligfield et al. (2007)
Augmented leads aVL, aVR and aVF
For aVR, aVL and aVF, the positive electrode is on the right arm, left arm and left leg respectively. The reference is the average of the other two measuring limb electrodes. Kligfield et al. (2007)
The angle of each frontal lead
In the conventional frontal-plane diagram, I is 0°, II +60°, III +120°, aVF +90°, aVL −30° and aVR −150°. Ashley and Niebauer (2004)
References
- Kligfield et al. (2007). Recommendations for the standardization and interpretation of the electrocardiogram, Part I. JACC 49, 1109–1127.
- Ashley and Niebauer (2004). Conquering the ECG. In Cardiology Explained, chapter 3.