ECG electrode placement errors and their effects

See how misplaced chest electrodes and reversed limb electrodes can change the appearance of an ECG tracing.

V1 (C1) sits immediately to the patient’s right of the sternum in the fourth intercostal space.
A word on 'lead': electrode, lead wire, lead

An electrode is a contact on the body. A lead is a measurement of the voltage difference between a positive input and a reference. The reference can combine signals from more than one electrode. Ashley and Niebauer (2004)

How often chest leads are misplaced

A study of clinicians and technicians found variation in where participants marked the precordial positions. V1 and V2 were frequently marked too high. The findings describe the study participants. Rajaganeshan et al. (2008, online 2007)

What a misplaced lead can imitate

Incorrect electrode position or a swapped wire can alter the tracing and mimic disease. For example, V1 and V2 placed too high can produce a pattern that resembles septal infarction. Harrigan, Chan and Brady (2012), Ilg and Lehmann (2012)

Never guess the intercostal space

Palpate the sternal angle to find the second rib. Count the spaces below the ribs to locate the fourth intercostal space on each side of the sternum. Kligfield et al. (2007)

Upper limb lead reversal

Example limb electrode positions near the wrists and ankles.
Example limb electrode positions near the wrists and ankles. (Figure 4 of the placement guide.)Illustration from the original 3D ECG Leads placement guide.

Interchanging the right-arm and left-arm wires changes the limb-lead pattern. Check the labelled connections when a recording appears inconsistent. Harrigan, Chan and Brady (2012)

Modified limb positions are a known source of false change

Right Leg Electrode Placement: The Right Leg Electrode RL(N) is placed above the right ankle if possible. It can be placed on the right thigh, if done so the left leg electrode should also be placed on the left thigh rather than the left lower leg. The leg should be resting on a surface to avoid muscle tension.
Right Leg Electrode Placement.Screenshot from the 3D ECG Leads app.

Moving limb electrodes onto the torso can change the recorded waveforms. Document a modified placement on the recording so it can be taken into account when tracings are compared. Jowett et al. (2005)

References

  • Kligfield et al. (2007). Recommendations for the standardization and interpretation of the electrocardiogram, Part I. JACC 49, 1109–1127.
  • Jowett et al. (2005). Modified electrode placement must be recorded when performing 12-lead electrocardiograms. Postgraduate Medical Journal 81, 122–125.
  • Harrigan, Chan and Brady (2012). Electrocardiographic electrode misplacement, misconnection, and artifact. Journal of Emergency Medicine 43, 1038–1044.
  • Ilg and Lehmann (2012). Importance of recognizing pseudo-septal infarction due to electrocardiographic lead misplacement. American Journal of Medicine 125, 23–27.
  • Rajaganeshan et al. (2008, online 2007). Accuracy in ECG lead placement among technicians, nurses, general physicians and cardiologists. International Journal of Clinical Practice 62, 65–70.
  • Ashley and Niebauer (2004). Conquering the ECG. In Cardiology Explained, chapter 3.