A word on 'lead': electrode, lead wire, lead
An electrode is a contact on the body. A lead is a measurement of the voltage difference between a positive input and a reference. The reference can combine signals from more than one electrode. Ashley and Niebauer (2004)
Where the six chest leads go

The positions below are described from the patient’s perspective. Kligfield et al. (2007)
- V1 (C1) sits immediately to the patient’s right of the sternum in the fourth intercostal space.
- V2 (C2) sits immediately to the patient’s left of the sternum in the fourth intercostal space.
- Place V4 (C4) in the fifth intercostal space on the patient’s left midclavicular line.
- Place V3 (C3) halfway along the line joining V2 and V4.
- Place V5 (C5) on the left anterior axillary line, level with V4.
- Place V6 (C6) on the left midaxillary line at the horizontal level of V4 and V5.
Why the fourth intercostal space must be found, not guessed

V1 and V2 have a defined recording position on either side of the sternum. Finding the fourth intercostal space gives you a repeatable landmark. Kligfield et al. (2007), Ilg and Lehmann (2012)
- An electrode one space too high can look plausible on the chest. High placement of V1 and V2 can produce a pattern resembling a septal infarct. Ilg and Lehmann (2012)
- Find the second rib at the sternal angle, then count down to the fourth space. Place V1 at the right sternal border and V2 at the left. Kligfield et al. (2007)
Sternum and jugular notch

The jugular notch is at the top of the sternum. The sternal angle lies lower down, at the junction of the manubrium and the body of the sternum. Kligfield et al. (2007)
Intercostal spaces and the sternal angle
Palpate the sternal angle to find the second rib. Count the spaces below the ribs to locate the fourth intercostal space on each side of the sternum. Kligfield et al. (2007)
Mid-clavicular and axillary lines


The midclavicular line runs vertically through the midpoint of the clavicle. V5 and V6 sit farther round the left chest at the same horizontal level as V4. Kligfield et al. (2007)
V1: step by step

V1 (C1) sits immediately to the patient’s right of the sternum in the fourth intercostal space. Kligfield et al. (2007)
- Step 1. Palpate the jugular notch.
- Step 2. Slide your fingers down until you feel a ridge (the sternal angle of Louis).
- Step 3. Move your hand and descend into the 2nd right intercostal space.
- Step 4. Move your hand down one rib to the 3rd right intercostal space.
- Step 5. Move your hand down one more rib to the 4th right intercostal space.
- Step 6. Place ECG chest lead V1(C1) in the 4th right intercostal space on the right border of the sternum.
V2: step by step

V2 (C2) sits immediately to the patient’s left of the sternum in the fourth intercostal space. Kligfield et al. (2007)
- Step 1. Palpate the jugular notch.
- Step 2. Slide your fingers down until you feel a ridge (the sternal angle of Louis).
- Step 3. Move your hand and descend into the 2nd left intercostal space.
- Step 4. Move your hand down one rib to the 3rd left intercostal space.
- Step 5. Move your hand down one more rib into the 4th left intercostal space.
- Step 6. Place ECG chest lead V2(C2) in the 4th left intercostal space on the left border of the sternum.
V4: step by step

Place V4 (C4) in the fifth intercostal space on the patient’s left midclavicular line. Kligfield et al. (2007)
- Step 1. Locate ECG chest lead V2(C2) in the 4th left intercostal space.
- Step 2. Descend one rib into the 5th left intercostal space.
- Step 3. Slide your hand along the 5th left intercostal space and stop at the left mid-clavicular line.
- Step 4. Place electrode V4(C4) on the left mid-clavicular line, in the 5th left intercostal space.
V3: step by step

Place V3 (C3) halfway along the line joining V2 and V4. Kligfield et al. (2007)
- Step 1. Locate ECG chest lead V2(C2) in the 4th left intercostal space.
- Step 2. Locate ECG chest lead V4(C4) in the 5th left intercostal space on the mid-clavicular line.
- Step 3. Slide your fingers and position them midway between V2(C2) and V4(C4).
- Step 4. Place ECG chest lead V3(C3) midway between V2(C2) and V4(C4).
V5: step by step

Place V5 (C5) on the left anterior axillary line, level with V4. Kligfield et al. (2007)
- Step 1. Locate ECG chest lead V4(C4) in the 5th left intercostal space on the mid-clavicular line.
- Step 2. Move your hand on a horizontal plane with ECG chest lead V4(C4) and stop on the left anterior axillary line.
- Step 3. Place ECG chest lead V5(C5) here, on the anterior axillary line, on a horizontal plane with ECG chest lead V4(C4).
V6: step by step

Place V6 (C6) on the left midaxillary line at the horizontal level of V4 and V5. Kligfield et al. (2007)
- Step 1. Locate ECG chest lead V5(C5) on the anterior axillary line, next to ECG chest lead V4(C4).
- Step 2. Move your hand on a horizontal plane with ECG chest lead V5(C5) and stop on the left mid-axillary line.
- Step 3. Place ECG chest lead V6(C6) here, on the left mid-axillary line, on a horizontal plane with ECG chest lead V4(C4) and ECG chest lead V5(C5).
References
- Kligfield et al. (2007). Recommendations for the standardization and interpretation of the electrocardiogram, Part I. JACC 49, 1109–1127.
- Ilg and Lehmann (2012). Importance of recognizing pseudo-septal infarction due to electrocardiographic lead misplacement. American Journal of Medicine 125, 23–27.
- Ashley and Niebauer (2004). Conquering the ECG. In Cardiology Explained, chapter 3.